Ультразвуковой дефектоскоп

Ультразвуковой дефектоскоп SD-UFD681

Модель: SD-UFD681

0–15,000 mm · 5.7″ IPS display · 500 channels · auto-calibration · PDF reports

  • Диапазон 0–15,000 mm (steel); transmission to 30 m
  • Дисплей 5.7″ IPS, 640×480, 60 Hz
  • Каналы / отчёты 500 / 1000
  • Динамический диапазон ≥100 dB
  • Отчёт PDF
Написать нам

Мин. заказ 1 шт. · Срок 5 рабочих дней · Ответ в течение 8 часов · Возможен OEM / ODM · Гарантия 2 года

WhatsApp QR code: scan to chat with SeiDiem
WhatsApp
WeChat QR code: scan to add SeiDiem on WeChat
WeChat

Технические характеристики приведены на английском языке для точности числовых значений. Для получения предложения или технического подтверждения на русском языке свяжитесь с нами напрямую.

The SD-UFD681 is a digital portable ultrasonic flaw detector designed for fast, damage-free detection, location and evaluation of internal defects (cracks, inclusions, pores) in welds and materials. The 5.7-inch IPS display offers a wide viewing angle with a 60 Hz refresh rate, and the ergonomic housing is water- and oil-resistant.

Auto-calibration covers probe zero offset, probe angle (K value) and material velocity. Up to 500 test channels and 1000 defect reports can be stored, and inspection results are exported as PDF reports.

Features

  • Ergonomic, waterproof and oil-proof housing with rotating damping bracket and rubber sheath
  • 5.7-inch IPS high-brightness display, no viewing angle, 640×480, 60 Hz refresh
  • Auto-calibration of probe zero offset, probe angle (K value) and material velocity
  • Real-time screenshots and PDF flaw-detection reports to USB drive
  • Peak-hold and edge trigger modes; echo gate and loss-of-echo gate
  • 500 test channels and 1000 defect reports storage
  • Straight, angle, dual-crystal and through-transmission probe support

Technical Data

Item Technical Parameters
Measuring Range 0–15000mm (at steel velocity); transmission up to 30 m
Material Velocity 100~20000m/s
Operational Stability Signal amplitude variation ≤±2%; signal time-base offset ≤±0.5%
 
Probe Frequency and Amplifier Bandwidth Narrowband Wideband
  1.0M(0.41-2.1)2.5M(1.4-3.6)5.0M(3.7-6.3)10M(8.2-11.8)15M(13-17) 1.0M(0.43-2.3)2.5M(1.0-4.0)5.0M(3.0-7.0)10M(7.5-12.5)15M(12.5-17.5)
Blind Zone After Transmit Pulse ≤9μs
Available Dynamic Range ≥100dB
Attenuator Error ≤±1.0dB(20dB span)≤±2.0dB(60dB span)
Vertical Amplitude Linear Error ≤2%
Horizontal Time-BaseLinear Error ≤0.5%
Repetition requencyand Accuracy ≤±5%(20~2000Hz)
Equivalent Input Noise ≤30 nV√Hz
Number of Testing Channels 500
Number of Defect Reports 1000
Testing Report Type PDF testing reports
Display Screen and Resolution 5.7-inch IPS high-brightness wide-viewing-angle display; 60 Hz refresh rate; 640×480
Digital Rejection 0–90% (does not affect linearity and gain)
Probe Types Straight probe, angle probe, dual-crystal probe, through-transmission probe
Gates Echo gate, loss-of-echo gate; single-gate reading, dual-gate reading
Trigger Mode Peak trigger, edge trigger
Alarm Functions Echo alarm, loss-of-echo alarm, buzzer alarm, LED alarm
Power Supply DC 9V; lithium battery (continuous operation ≥12 hours)Charger: AC Mains 100-240 VAC, 50-60 Hz
Auxiliary Functions Screenshot, recording and playback, B-scan, automatic DAC/AVG curve generation,AWS intelligent acceptance rating, data analysis software
Advanced Functions Weld diagram; probe frequency detection
Video Recording Capacity Internal storage: 10 video files(up to 5 minutes each, 30 frames per second real-time waveform);External USB: unlimited files/duration
Dimensions 230×157×52 mm
Weight 1.13 kg
Operating Temperature -30–55°C
Relative Humidity 20–95% RH
Protection Level IP62
Testing Standards Compliant with EN12668-1 and other industry standards
Pulse Type Negative pulse wave; transmit pulse voltage: -100V–-500V (selectable in grades)
Pulse Width 30nS ~ 510nS(continuously adjustable to match different probe frequencies)
Pulse Shift -7.5 μs–3000 μs
Pulse RepetitionFrequency ranges 20 Hz to 2000 Hz
Probe Damping 50Ω, 150Ω, 300Ω, 400Ω (optional, to meet sensitivity and resolution requirements)
Probe Zero Value 0 μs–999.9 μs
Sampling Rate 12-bit AD sampling, 200 Msps, high-fidelity waveform
Detection Modes Positive half-wave; negative half-wave; full wave; RF wave
Maximum Gain 110 dB (step: 0.1dB, 1dB, 2dB, 6dB)
Communication Interfaces USB/RS232
Probe Interfaces BNC/LEMO
 
System Linearity Vertical: 1.5% FSH, Amplifier Accuracy +/-1 dB. Horizontal: +/-0.1% FSW
Sensitivity Leavings more 50dB(Deep 200mmФ2 flat-bottomed hole)
Dynamic range more 36dB
Pulse Displacement more 32dB
Resolving Power more 42dB
Noise less than15%
Display Delay -20 to 3400 µs
Probe Delay/Zero Offset 0 to 99.99µs
Test Modes Pulse echo, dual element and thru-transmission
Gate Monitors Two independent gates controllable over entire sweep range
Rectification Positive half wave, negative half wave, full wave, RF
Reject (suppression) 0 to 80% full screen height
Units Inch or millimeter
 
Note All indicators measured under: probe frequency 2.5 MHz, detection mode full wave.
 
 