Ultrasonic Thickness Gauge Shows Nothing? Check These Five Points

Thickness Gauges 5 min read

The same gauge worked yesterday and shows nothing today. In most cases the instrument is fine - work through these five checks before sending it away.

1. No echo at all

Start with coupling. Rust, paint, scale or a rough surface traps a layer of air between probe and part, and the sound never gets in. Grind or file to bare metal, then apply couplant.

Then check angle: if the front and back surfaces are not parallel, the back-wall echo scatters away. On curved parts use a small-diameter probe – a flat probe on a curved surface touches only at a point or along a line.

Castings and austenitic steel deserve special mention: coarse grains scatter heavily and the echo drowns. Switch to a lower-frequency, coarse-grain probe.

2. Readings jump or disappear intermittently

Usually a worn probe face. The acrylic wear layer clouds with use and sensitivity falls. Resurface with 500-grit paper while keeping it parallel; if it is still unstable, replace the probe.

A back wall with heavy pitting does the same: the pits attenuate the beam, readings wander and sometimes vanish.

3. Readings too high

Three common causes. Internal deposit whose acoustic impedance is close to the steel – the gauge shows wall plus deposit. A dense oxide or paint layer, because sound travels at a different speed in each material. Internal defects such as inclusions, where the reading often lands near 70 percent of nominal thickness and a flaw detector is needed to locate the defect.

4. Readings too low or unstable

Wrong velocity is the most common: calibrating on steel and then measuring aluminium guarantees an error. Confirm the material or reverse-measure velocity on a standard.

Hot work also misreads: sound speed in solids falls by roughly one percent per 100 degrees Celsius. In-service hot equipment needs a 300-600 degrees probe; a standard probe must not be used.

5. Unit dies mid-measurement

Replace the battery first. Then check the probe connector for oxide or a loose fit – connector faults are a large share of field failures. Couplant that has seeped into the housing causes key failure, so wipe the instrument down after every session.

Summary

Surface, coupling, probe, velocity, power – in that order. Nine out of ten cases stop at the first step.